 |
| Volume 3, Number 2, Article 5, Pages 171-182 |
doi:10.1167/3.2.5 |
http://journalofvision.org/3/2/5/ |
ISSN 1534-7362 |
Topography of the chromatic pattern-onset VEP
Christina Gerth |
Department of Ophthalmology and Section of Neurobiology, Physiology and Behavior,
University of California, Davis, CA, USA |
|
Peter B. Delahunt |
Department of Ophthalmology and Section of Neurobiology, Physiology and Behavior,
University of California, Davis, CA, USA |
|
Michael A. Crognale |
Departments of Psychology & Biomedical Engineering Program, University of Nevada, Reno, NV, USA |
|
John S. Werner |
Department of Ophthalmology and Section of Neurobiology, Physiology and Behavior,
University of California, Davis, CA, USA |
|
Abstract
The chromatic pattern-onset VEP has been used successfully as a sensitive and objective technique to determine congenital and acquired color vision deficiency. It also has been applied to characterize development, maturation and aging of the chromatic visual pathways. Here we determine the topographic components of the full-field VEP using the multifocal technique. Recordings were made with the VERISTM system that extracts topographic VEPs using a pseudorandom stimulus sequence. Chromatic pattern stimuli were presented in an onset-offset temporal sequence, with colors modulated along different axes in the MBDKL color space. Additional experiments were conducted to verify the S-cone axis for each observer and that our chromatic stimuli were close to isoluminant at different field locations. Our data show reliable and robust chromatic onset VEP responses for multiple retinal areas that conform to pattern-onset full-field VEP waveform characteristics. For stimuli with chromatic contributions, pattern-onsets produced reliable and consistent waveforms whereas for stimuli with large luminance contributions pattern-reversal stimuli were superior. Our method for recording chromatic multifocal pattern-onset VEPs holds promise for clinical application to detect and monitor early retinal and optic nerve changes related to aging and disease.
 |
|
History
Received September 25, 2002; published April 3, 2003
Citation
Gerth, C., Delahunt, P. B., Crognale, M. A., & Werner, J. S. (2003). Topography of the chromatic pattern-onset VEP.
Journal of Vision, 3(2):5, 171-182,
http://journalofvision.org/3/2/5/,
doi:10.1167/3.2.5.
Keywords
multifocal, VEP, chromatic axes, color
for related articles by these authors
for papers that cite this paper |
Visual evoked potential (VEP) recordings are generally
accepted as a sensitive and objective test of the integrity of the visual system
(e.g. Harding, Odom, Spileers & Spekreijse,
1996). To investigate the focal contribution to the full-field VEP, Baseler
et al. ( Baseler, Sutter, Klein & Carney,
1994) utilized the multifocal recording technique developed by Sutter ( Sutter, 1991; Sutter & Tran, 1992) to record VEPs
simultaneously from multiple locations in the visual field. This multifocal
technique permits recording and analysis of multiple focal retinal (multifocal
electroretinogram) or cortical (multifocal VEP or mfVEP) responses. Several
groups have already applied the multifocal technique to record pattern-reversal
mfVEPs in normal subjects and in patients with diseases such as glaucoma ( Klistorner, Graham, Grigg & Billson,
1998; Klistorner & Graham, 2000;
Hood et al, 2000; Hasegawa & Abe, 2001) and amblyopia ( Yu, Brown & Edwards, 1998). Detecting response
changes by the mfVEP technique prior to visual field losses in glaucoma would
provide important information about the onset and progression of the disease
(see Goldberg, Graham and Klistorner,
2002).
It is well known that an S-cone pathway is affected
early in diseases such as glaucoma, diabetes and retinitis pigmentosa ( Sandberg & Berson, 1977; Adams, Rodic, Husted & Stamper, 1982; Greenstein, Hood, Ritch, Steinberger &
Carr, 1989). Chromatic cortical responses can be elicited with the
pattern-onset VEP using low spatial frequency isoluminant stimuli ( Berninger, Arden, Hogg & Frumkes, 1989;
Murray, Parry, Carden & Kulikowski,
1987; Rabin, Switkes, Crognale, Schneck &
Adams, 1994). Crognale et al. ( Crognale et
al., 1993 1993) have shown that chromatic onset VEPs are
useful in detecting congenital and acquired color vision deficiency. To be able
to determine early and localized abnormalities in the chromatic pathways would
be a dramatic improvement in objectively detecting and monitoring retinal and
optic nerve diseases. MfVEPs in the chromatic pattern-onset mode have not
previously been described in the literature.
The purpose of this paper is to determine the
topographic components of the full-field chromatic onset VEP by recording
multifocal responses in normal subjects. Using isoluminant stimulus patterns we
were able to record along the tritan (S cone) and L-M axes to confirm the
response characteristics described for chromatic onset full-field VEPs ( Murray, Parry, Carden & Kulikowski, 1987,
Berninger, Arden, Hogg & Frumkes,
1989; Rabin, Switkes, Crognale, Schneck &
Adams, 1994). We also obtained responses to isochromatic stimuli that varied
in luminance. Additionally, several experiments were conducted to verify
isoluminance and S-cone isolation for individual observers.
MfVEPs 1 were
obtained from five normal subjects ages 23 to 38 years. The presence of retinal
disease and abnormal ocular media in the tested eye were ruled out by ocular
examination including visual acuity, slit lamp examination, intraocular pressure
and direct and indirect ophthalmoscopy. Color stereo fundus photographs of the
macula and optic disc (ETDRS Fields 1 and 2) were evaluated by a retinal
specialist using a stereo viewer. The retinae of all subjects were found to have
no vascular, retinal, choroidal or optic nerve findings known to disrupt visual
function. Intraocular pressure was ≤ 22 mm Hg. All subjects demonstrated a
corrected Snellen acuity of ≥ 20/20 in the tested eye, as well as normal
color vision when tested with the Neitz anomaloscope, the HRR pseudoisochromatic
plates and the Farnsworth F-2 plate. Written informed consent was obtained
following the Tenets of Helsinki, and with approval of the Office of Human
Research Protection of the University of California, Davis, School of Medicine.
Equating Perceptual Contrast
Before any mfVEP measurements were made, we first
selected stimulus contrasts along the three color dimensions to be tested
(tritan, L-M, and luminance axes). Following Switkes and Crognale ( 1999) we used a psychophysical approach to
obtain perceptually equated stimuli for suprathreshold contrast for our
experiments.
The stimuli were presented on the same monitor (Sony
Trinitron 20”) used for the mfVEP experiment driven by a Macintosh G4
computer with an 8 bit IMS Twin Turbo graphics card. The stimulus arrangement
was equivalent to the stimulus pattern used by the VERIS TM system
except the hexagons were not scaled. For each image, the 19 hexagons were
presented simultaneously with a fixation point placed at the center. Illuminant
C was chosen as the white point (CIE x,y = 0.310,0.316) and was used as the
stimulus background. The test stimuli were modulated from the white point along
the three cardinal axes of the MBDKL color space ( MacLeod & Boynton, 1979; Derrington, Krauskopf & Lennie, 1984).
The chromatic stimulus patterns (tritan and L-M) were isoluminant across the
whole image at 37 cd m -2, and the luminance stimulus pattern had a
mean luminance value of 37 cd m -2.
The experimental software was written in MATLAB ( http://www.mathworks.com/) using the
Psychophysics Toolbox extensions ( Brainard,
1997; Pelli, 1997). The monitor was
calibrated using a Minolta colorimeter (CS 100 Chroma Meter) and procedures set
out in Brainard et al. ( Brainard, Peli &
Robson, 2002). The CIE 1931 color-matching functions and Smith-Pokorny cone
fundamentals ( Smith & Pokorny, 1975) were
used to convert between the measured monitor RGB outputs and cone
stimulation.
Using the calibration data, we calculated the maximum
distance from the white point in each cone direction using L*a*b* space (an
approximately uniform perceptual color space). The least of these values (87.5
DE units) was used as the furthest step for each of the chromatic stimuli. Ten
equally spaced step sizes (8.75 DE units) were calculated from the white point
to the furthest step (see Figure 1).
For the luminance contrast settings, the same chromaticity coordinates as the
white point were used, but consisted of 10 equally spaced luminance contrast
values above and below the mean luminance
value. Figure 1 . Chromatic axes and step sizes are shown in both CIE a*b*
and xy color coordinates. The center point is Illuminant C.
Figure 2 . Data are shown (blue dots) for one subject (ABR).
Luminance (top panel) and L-M (bottom panel) stimuli of varying contrasts were
compared to the tritan standard. A psychometric function was fitted to the data
(green curves) to estimate the PSE (red dashed lines).
An S-cone contrast setting was chosen as the standard
and the L-M and luminance stimuli were tested against it. This S-cone setting
was chosen in a pilot experiment so that there were approximately equal step
sizes above and below this setting for the other two dimensions. A 2AFC constant
stimulus procedure was used. For each stimulus pair, the standard was shown
against either an L-M or luminance stimulus. The presentation order was
randomized, as was the order of the contrast step-size of the test stimuli. Each
stimulus pair was presented sequentially and the observer indicated which
interval contained the highest contrast by pressing either ‘1’ or
‘2’ on a number keyboard. Presentation time was 1 s for each
stimulus pattern, with a 0.2 s interval between the pair. Both sets of stimuli
(L-M and luminance) were tested in the same experiment monocularly. Each
contrast step was tested five times requiring a total of 100 trials (2
dimensions x 10 step sizes x 5 presentations).
The data were fitted with a cumulative normal
psychometric function and the point of subjective equality (PSE) was the
contrast at the 50% point (see Figure 2
for example). The values at these points were converted to RGB input values for
the VERIS TM system using the calibration data. A separate set of
values was calculated for each subject. The cone values and Michelson contrasts
are shown in Table 1. Note that the values are
similar across
subjects. Table 1 . Perceptually Equated Cone Stimulation Values and Michelson
Contrasts.
|
|
S+
|
S-
|
Cont.
|
L+M-
|
L-M+
|
Cont.
|
Lum+
|
Lum-
|
Cont.
|
|
ABR
|
L
|
0.0762
|
0.0757
|
0%
|
0.0790
|
0.0729
|
4%
|
0.0813
|
0.0706
|
7%
|
|
|
M
|
0.0645
|
0.0642
|
0%
|
0.0593
|
0.0692
|
-8%
|
0.0689
|
0.0598
|
7%
|
|
S
|
0.0730
|
0.0288
|
43%
|
0.0490
|
0.0490
|
0%
|
0.0524
|
0.0455
|
7%
|
|
CIS
|
L
|
0.0762
|
0.0757
|
0%
|
0.0790
|
0.0730
|
4%
|
0.0812
|
0.0707
|
7%
|
|
|
M
|
0.0645
|
0.0642
|
0%
|
0.0594
|
0.0691
|
-8%
|
0.0688
|
0.0599
|
7%
|
|
|
S
|
0.0730
|
0.0288
|
43%
|
0.0490
|
0.0490
|
0%
|
0.0524
|
0.0456
|
7%
|
|
CG
|
L
|
0.0762
|
0.0757
|
0%
|
0.0791
|
0.0728
|
4%
|
0.0814
|
0.0705
|
7%
|
|
|
M
|
0.0645
|
0.0642
|
0%
|
0.0592
|
0.0693
|
-8%
|
0.0690
|
0.0597
|
7%
|
|
|
S
|
0.0730
|
0.0288
|
43%
|
0.0490
|
0.0490
|
0%
|
0.0525
|
0.0454
|
7%
|
|
PBD
|
L
|
0.0762
|
0.0757
|
0%
|
0.0790
|
0.0729
|
4%
|
0.0812
|
0.0707
|
7%
|
|
|
M
|
0.0645
|
0.0642
|
0%
|
0.0594
|
0.0692
|
-8%
|
0.0688
|
0.0599
|
7%
|
|
|
S
|
0.0730
|
0.0288
|
43%
|
0.0490
|
0.0490
|
0%
|
0.0524
|
0.0456
|
7%
|
|
SMG
|
L
|
0.0762
|
0.0757
|
0%
|
0.0785
|
0.0734
|
3%
|
0.0805
|
0.0714
|
6%
|
|
|
M
|
0.0645
|
0.0642
|
0%
|
0.0601
|
0.0685
|
-6%
|
0.0682
|
0.0605
|
6%
|
|
|
S
|
0.0730
|
0.0288
|
43%
|
0.0490
|
0.0490
|
0%
|
0.0519
|
0.0461
|
6%
|
Values for each of the five subjects are listed.
Note that the S-cone contrast was fixed and the contrasts along the other two
axes (L-M and luminance) were equated to it psychophysically.
MfVEP Stimulus and Procedure
The stimulus array was produced and displayed on a CRT
using the VERIS TM 4.8 multifocal system. This system has proven to be
a useful tool for multifocal recording particularly in clinical environments
(see http://www.cephalon.dk/ for further
information). It is also useful for experimental work but for our purposes there
are two limitations worth noting. First, the choice of stimulus patterns is
currently limited. The system does offer a cortically scaled pattern (dartboard)
but it contains many edges (especially in the central area) that produce a
substantial level of high spatial frequency components. For our purposes it was
more important to reduce high spatial frequency components and the
‘triangle pattern’ was therefore chosen. Note that further
VERIS TM software will provide an option to import customized stimulus
patterns (Erich Sutter, personal communication, December 2002).
Second, the system includes a customized graphics card
that currently allows only 101 values per gun. Although this resolution is
relatively low, a control experiment (see next section) suggests that the
resolution level was sufficient for our purposes.
The pattern we used contained 19 scaled hexagons
(although not cortically scaled) and each hexagon consisted of 24 triangles (see
Figure
3 and 4). The stimuli were viewed from a
distance of 60 cm and had a retinal subtense of 14 degrees in radius with a
central hexagon over 2.5 degrees, a second ring (containing 6 hexagons) from 2.5
– 8 degrees and an outermost ring (12 hexagons) from 8 – 14 degrees.
The stimulus was presented on a 20” SONY Trinitron color monitor (frame
rate 75 Hz). Figure 3 . The
VERIS TM triangle pattern
is shown. The red rings are superimposed to illustrate the extent of retinal
stimulation (in degrees radius).
Figure 4 . This movie shows the L-M stimulus pattern used in
Experiment 1.
MfVEPs were recorded monocularly (same eye that was
tested in the color contrast experiment) with silver-silver chloride electrodes
placed at O z (active), F z (reference) and C z
(ground) following the International 10/20 system ( Harding, Odom, Spileers & Spekreijse,
1996). The scalp-electrode impedance was ≤ 6 kΩ. The subjects
were instructed to fixate the target (black cross) in an alert condition. Each
recording was divided into 32 segments, each 51.18 sec long with a resulting
total recording time of 27 min 18 sec. The continuously recorded mfVEP was
amplified (10 5), band-pass filtered at 1-100 Hz (GRASS preamplifier
CP511) and sampled at 1200 Hz (every 0.83 ms). The m-sequence was set to
2 12–1. First-order kernel responses were extracted and analyzed
using VERIS TM 4.8.
The nomenclature
(C I/C II/C III) recommended by the ISCEV Standard
for Visual Evoked Potentials 1995 ( Harding et
al., 1996) was used for the response analysis (see Figure 5). In our analysis we paid
particular attention to C I and C II because typically
C I is larger for isochromatic luminance modulation and C II
is larger for isoluminant chromatic modulation. In our experiment,
C III covaried with C II and therefore did not provide any
further information. Thus, we have restricted our discussion to C I
and C II.
Figure 5 . The three main components of a pattern-onset VEP waveform
are shown. (This waveform is a response to one of our experimental conditions
that varied in both luminance and chromaticity.)
Verification of Tritan Axis Stimulation
Before proceeding with the mfVEP recordings, we wanted
to verify that our stimuli produced the desired pathway stimulation. There are a
number of possible reasons why our stimulus patterns might not do this. As
mentioned above, the VERIS TM system has relatively low color
resolution (101 values per gun). Also, the cone stimulation values we used are
based on Smith-Pokorny’s ( 1975) cone
fundamentals while the actual values for individual observers with normal color
vision will vary slightly. Finally, the calibration procedure assumes monitor
spatial homogeneity and phosphor independence and small violations may also
introduce some degree of error.
We tested whether our tritan stimulus matched the
tritan axis for each of the observers used in the main experiment with a
transient tritanopia paradigm ( Mollon &
Polden, 1977; Crognale et al., 1995).
We asked observers to rate the stimulus contrast before and after adaptation to
a bright 580 nm field. Following the extinction of this adapting field,
thresholds for lights detected by a short-wavelength cone pathway are elevated
due to polarization of a cone-opponent site of adaptation. If the stimulus
pattern were sufficiently close to the observer’s tritan axis, the
perceived contrast of the pattern should be reduced after adaptation.
Subjects viewed, with the same eye to be tested in the
mfVEP experiment, a 580 nm, 16° diameter, 10.5 log quanta sec-1
deg-2 adapting background in Maxwellian view for 30 sec. They then
immediately inspected one of the three test patterns used in the mfVEP
Experiment. Ratings of stimulus contrast were made for the adapted and unadapted
eyes before and after adaptation.
As expected, there was no intra-ocular transfer of
adaptation as evidenced by similar ratings before and after adaptation in the
unadapted eye. The tritan stimulus was essentially invisible to all subjects
following adaptation, while the L-M axis and luminance stimulus were essentially
unchanged, if not slightly enhanced in their perceived contrast. Table 2 shows the individual ratings for
each stimulus pattern. This experiment demonstrates that in all subjects the
tritan stimulus used for the mfVEP recording was effective at isolating an
S-cone
pathway. Table
2 . Perceived Contrast Before and After Transient
Tritanopia.
|
Subject
|
Pattern
|
Right
eye
|
Left
eye
|
Right eye
|
Left
Eye
|
|
ABR
|
Tritan
|
10
|
10
|
10
|
0
|
|
|
L-M
|
10
|
9
|
10
|
10
|
|
|
Luminance
|
10
|
9
|
10
|
13
|
|
CIS
|
Tritan
|
10
|
9
|
9
|
0
|
|
|
L-M
|
10
|
10
|
10
|
9
|
|
|
Luminance
|
10
|
10
|
10
|
12
|
|
CG
|
Tritan
|
10
|
10
|
10
|
0.5
|
|
|
L-M
|
10
|
10
|
10
|
10
|
|
|
Luminance
|
10
|
10
|
10
|
11
|
|
PBD
|
Tritan
|
10
|
10
|
10
|
0
|
|
|
L-M
|
10
|
10
|
10
|
10
|
|
|
Luminance
|
10
|
10
|
10
|
10
|
|
SMG
|
Tritan
|
10
|
10
|
10
|
0
|
|
|
L-M
|
10
|
9
|
10
|
9
|
|
|
Luminance
|
10
|
10
|
10
|
10
|
After adaptation, there is essentially no perceived
tritan contrast (highlighted in blue).
Experiment 1: Chromatic Pattern-Onset mfVEP
MfVEP responses were recorded when subjects were
presented with perceptually equated contrast pattern-onset stimuli. Contrast was
modulated along three axes (luminance, L-M, and tritan) as described in detail
above.
Figure 6 shows mfVEP waveforms for two subjects (CG and CIS) and the average for
all 5 subjects: the S axis responses (blue) are superimposed on either the L-M
axis responses (red; left panel) or the luminance responses (black; right
panel). For both chromatic and luminance modulation, the second response
component (C II) is faster and smaller with increasing eccentricity.
The responses are largely dominated by the fovea, which is demonstrated by
robust waveforms from the central area. Peripheral responses suggest a lower
signal-to-noise ratio with smaller response amplitudes than central responses. 2 In the majority of the recordings, waveforms
in the superior and inferior field do not demonstrate an obvious change in
polarity as described for the pattern-reversal mfVEP (e.g. Baseler, Sutter, Klein & Carney, 1994; Baseler & Sutter, 1997; Yu & Brown, 1997). The use of single-channel
recording and the electrode positioning (see Klistorner, Graham, Grigg & Billson,
1998 and Hood, Zhang, Hong & Chen,
2002) might contribute to the failure to observe polarity changes in our
records.
Figure 6 . Response
waveforms are shown for two subjects and for the mean of all 5 subjects. The
waveforms in the left column are for tritan (blue) and L-M (red) responses. The
right column shows the tritan (blue) and luminance (black) responses.
Tritan
responses: In agreement with previous studies on chromatic pattern-onset
VEPs (e.g. Rabin et al., 1994), the responses
along the S axis are characterized by a negativity (C II) after 100 to
160 ms. In all subjects the slowest C II latencies were
found in the central/foveal response (subjects: CG: 148 ms; CIS: 132
ms; ABR: 134 ms; SMG: 129 ms; PBD: 152 ms) compared to the other 18 areas
tested. Most of the responses did not exhibit a well-formed positive
(C I) component.
L-M axis
responses: The responses to L-M modulation are similar in magnitude and
shape to the tritan responses. C II latencies for the L-M responses
were found to be slightly faster than the tritan responses as demonstrated in
the upper panel in Figure
6and as reported previously for the full-field VEP (e.g. Rabin et al., 1994).
Luminance
responses: Responses to luminance modulation are characterized by smaller
C II components, which are less pronounced in the periphery compared
to chromatic responses. More often than in the chromatic responses there are
well-demarcated C I luminance response components. Based on previous
recordings of low spatial frequency pattern-onset VEPs ( Murray et al., 1987; Rabin et al., 1994), one might expect smaller
responses to luminance onset pattern stimuli. We assume that the present
luminance responses are larger due to
additional contributions from mechanisms sensitive to
high spatial frequencies in our patterns. We investigate this issue further in
Experiment
3.
Experiment 2: Chromatic Pattern-Reversal mfVEP
Full-field VEP studies have shown that for low spatial
frequency stimuli, pattern-onset stimulation is superior to the pattern-reversal
presentation for recording robust and reliable isoluminant chromatic responses
( Murray et al., 1987; Berninger et al., 1989; Rabin et al., 1994). We ask whether this
full-field response characteristic is valid for the multifocal stimulation
technique and repeated the previous stimulation conditions but with the reversal
presentation.
We recorded the tritan pattern-reversal mfVEP for one
of the subjects (PBD) with the same stimulus pattern, color settings and data
acquisition as we used for the pattern-onset stimulation. To accomplish the
reversal stimulation, one frame per m-step was used. The m-sequence was set to
2 12–1. The recording was divided into 2 segments, each 27.30
sec long with a resulting total recording time of 55 sec per run. The surround
was set to Illuminant C. Responses for the first slice of the second-order
kernel, which represent the interaction between two consecutive frames of the
monitor ( Baseler et al., 1994) were
extracted and analyzed using VERIS TM 4.8.
To directly compare the pattern-reversal and the
pattern-onset mfVEP responses, we repeated the pattern-onset test in the same
session and with the same electrode position. The onset stimulus used 6
‘on’ frames and 6 ‘off’ frames for each m-sequence step.
Pilot recordings using the temporal settings of 6 chromatic or luminance
stimulus frames as ‘on’ and 6 blank frames as ‘off’ did
not differ from the 6 ‘on’/24 ‘off’ in their waveform
characteristics. The recording was divided into 16 segments, each 41.23 sec long
with a resulting total recording time of 10 min 36 sec per run. Blank frames and
the surround were set to Illuminant C.
A mean luminance value of 30 cd . m -2
was used for all conditions in this control experiment. 3 Two runs of both the reversal and
onset conditions were performed in the same session so that a two-run signal to
noise ratio (2rSNR) analysis could be performed ( Zhang, Hood, Chen, Hong, 2002). The responses from
45 – 300 ms were used in the analysis (the responses in the first 45 ms of
the recordings are due mainly to cortical noise).
In Figure 7
individual waveforms for the pattern-onset mfVEP (top panel) and
pattern-reversal mfVEP (bottom panel) are shown. In contrast to the tritan
pattern-onset waveforms, the pattern-reversal responses are less distinct and
smaller in their amplitude. The pattern-reversal responses exhibit a small
negativity at around 100 to 120 ms having decreasing amplitudes with increasing
retinal eccentricity. The rather typical sharp negativity (C II),
which is evident for the tritan onset responses, is less distinct for the
pattern-reversal presentation. These findings parallel full-field VEP responses.
Figure 7 . Tritan responses are shown for both pattern-onset stimuli
(top panel) and pattern-reversal stimuli (bottom panel). The red and blue traces
show the responses from two different runs in the same session. The 2rSNR values
are shown above each set of responses.
The 2rSNR analysis results are shown above each set of
responses. Values of zero indicate no signal, and values above 1 have a high
probability of a signal being present ( Zhang et
al., 2002). The 2rSNR values for the onset responses are significantly
greater than for the reversal responses (t-test, p < .01).
The onset stimulus used 12 frames for each step of the
m-sequence, whereas the reversal stimulus used only 1 frame per step. Therefore
the recordings for the onset conditions are twelve times longer. The important
consideration was to equate the m-sequence for both types of stimulus
presentation so that the sampling rates were identical.
The results confirm that the pattern-onset stimulation
is better for studying chromatic mfVEP responses for both the full-field ( Murray et al., 1987; Rabin et al., 1994) and localized
contributions.
Experiment 3: Control for High Spatial Frequency Contributions
As mentioned previously, the choice of stimulus
patterns available with the VERISTM system is limited and the pattern
we chose contained sharply defined edges that introduced undesirable
high-spatial frequency components. To help determine the degree to which these
components affected our results, we conducted an additional experiment with a
lens placed in front of the observer’s eye to blur the stimulus.
One of the subject’s (CG) mfVEP recordings were
repeated for the luminance and the tritan stimulus on the same eye tested as in
the original experiment. The stimulus was blurred with a + 3.0 D trial lens, 4 which would be expected to shift the high
spatial frequency cut-off to
< 6 cycles per degree ( Westheimer, 1964).
In Figure 8
waveforms are superposed for ‘in focus’ (red) and
‘blurred’ (blue) recordings with the tritan stimulus (top panel) and
the luminance stimulus (bottom panel). Responses to both stimulus patterns are
affected, however, the luminance responses are affected to a larger degree. The
C I component seems to be less affected than the C II and
C III components when the stimulus is blurred. For the luminance
response, the C II and C III components are almost
extinguished under the blur condition. It is likely that for stimuli with even
lower spatial frequencies, the luminance response would be extinguished. For the
S stimulus the responses still exhibit well-demarcated C II and
C III components, which are smaller in amplitude and slower in peak
latency than in the ‘in focus’ response. Overall, the sharp
negativity (C II) of our ‘in focus’ luminance stimulus
appears to depend on the presence of high spatial frequency components. The
tritan responses are preserved with induced blur. This agrees with previous
reports using low spatial frequency stimuli where tritan responses are much more
pronounced than luminance responses (e.g. Rabin
et al., 1994). Figure 8 . The waveforms for in-focus (red traces) versus blurred
(blue traces) stimuli are shown for tritan responses (top panel) and luminance
responses (bottom panel).
Experiment 4: Control for Isoluminance
Previous research has shown that the C I
component is maximized for isochromatic luminance modulation and C II
for isoluminant chromatic modulation (e.g. Rabin et al., 1994). The purpose of this
experiment was to determine which components are generated by chromatic versus
luminance pathways in the mfVEP by varying the luminance and chromatic
components of those stimuli. To do this, the original chromatic axes were tilted
into the luminance plane by varying amounts to change the relative chromatic and
luminance modulation of the stimuli. MfVEPs were recorded with these new
stimuli.
Eight new test axes were created using a combination of
the chromatic and luminance axes used in the
General Methods/Perceptual Contrast
Experiment. The contribution of each component was based on the angle of
tilt. For example, for a +30 degree tilt, the axis consisted of 50% of the
original luminance axis and 88.6 % of the original chromatic axis (see Figure 9). The two original chromatic axes
were tilted into the luminance plane using four different angles (+/- 30 degs,
+/- 60 degs). Switkes and Crognale ( 1999)
showed that the PSE for one chromatic axis could predict the PSE on another. We
therefore estimated the new PSEs for the new axes using the results from the
perceptual contrast experiment (see General Methods section).
Figure 9 . The tilted axes are illustrated on the left. The stimulus
contrasts are illustrated on the right. (The colors are approximations only. In
the experiments, all stimuli were presented on a carefully calibrated
monitor.)
A mean luminance value of 30 cd . m-2
was used for this set of experiments. To avoid inter-session variability,
the original three conditions were repeated together with the eight new
conditions in one test session. The onset stimulus used 6 ‘on’
frames and 6 ‘off’ frames for each m-sequence step. The m-sequence
was set to 212–1 and the first-order kernel responses were
extracted and analyzed using VERISTM 4.8. Blank frames and the
surround were set to Illuminant C. All 11 tests were recorded on three of the
original five subjects (CG, CIS, PBD) within one test session.
If the chromatic stimuli are close to isoluminance, the
results from the tilted axes should fall between the chromatic and luminance
responses. Figure 10 shows the
responses from the central area for the tritan, luminance and tilted axis
responses for one subject. Note that the tilted responses do indeed fall between
the chromatic and luminance responses and that the +/- 30 degree responses are
closer to the chromatic response and the +/- 60 degree responses are closer to
the luminance response. This provides compelling evidence that the chromatic
axis is close to isoluminant. To summarize the responses, we measured the CII
response density using the scalar product method (see Sutter 1992) for each of the waveforms.
C II was chosen because it is large for chromatic modulation and low
for luminance modulation. Figure 11
shows the C II response densities for the tritan axis for the central
region for one observer
(CG). Figure 10 . The waveform responses to the tritan, luminance and
tilted axes are shown for the central region for one subject. Note that the
responses to the tilted axes are generally between the tritan and luminance
responses suggesting that the chromatic stimuli are close to isoluminant.
Figure 11 . These are the C II
response densities for the waveforms shown in Figure 10. The inverted-U shape indicates
that the chromatic stimuli are isoluminant.
An inverted-U shape suggests that the chromatic stimuli
were close to isoluminant (recall that the C II component largely
reflects inputs from chromatic pathways). This inverted-U shape was found in the
central, parafoveal and some of the outer regions (see Figure 12). Some of the outer regions had
weak responses but no location showed a luminance response contribution to the
C II component. This indicates that a pattern that is isoluminant for
the central region is close to isoluminant in the periphery. Luminance artifacts
due to retinal inhomogeneity do not seem to have much of an impact on the
peripheral results using the present procedure. The tritan responses are shown
because this axis is particularly vulnerable to violations of isoluminance (due
to retinal inhomogeneity of macular pigment density). The results for the L-M
axis were similar.
Figure 12 . These plots show the
C II response densities for one subject for
the tritan, luminance and tilted axes. Each subplot is in the same format as Figure 11. The inverted-U shape was found
in the central region, middle regions and some outer regions. The red rings
illustrate the radius (in degrees) of the stimulated retinal areas..
The VERIS TM system is a useful tool for
recording multifocal responses. It has, however, been designed mainly for
clinical applications and the software presently available has some limitations
for experimental use (in particular, relatively low color resolution and a
restricted choice of stimulus patterns). Despite these limitations, we were
successfully able to record chromatic onset mfVEPs at multiple field locations.
In agreement with previous research ( Murray et
al., 1987; Berninger et al., 1989; Rabin et al., 1994) we found that chromatic
responses are much larger using the pattern-onset rather than the
pattern-reversal mode of presentation. The chromatic onset responses are similar
in their waveform and latency to full-field responses reported previously ( Murray et al., 1987; Rabin et al., 1994). The responses obtained from
the two chromatic axes we used (tritan and L-M) are mainly dominated by the
fovea. Responses in the peripheral field are smaller and have a lower
signal-to-noise ratio.
With the transient tritanopia experiment we have
demonstrated that in all subjects the stimulus we used for mfVEP recording
essentially stimulates only an S-cone pathway for the entire test field.
Further, the tilted axes experiment showed that the C II response
amplitude peaked at the isoluminant point and was minimal at the luminance point
in the central and in more than 50% of the peripheral responses extending 14
degrees in radius. No obvious waveform polarity differences were observed
between the superior and inferior areas of the field. This is likely to be due
to the electrode placement and the one channel recording technique ( Klistomer et al, 1998; Hood, Zhang, Hong, & Chen, 2002).
The validity of large field chromatic VEPs and mfVEPs
depends on minimal disruption of isoluminance due to retinal inhomogeneity ( Kulikowski, Robson & McKeefry, 1996; Switkes, Crognale, Rabin, Schneck & Adams,
1996). Several authors (e.g. Rabin et al.,
1994; Porciatti & Sartucci, 1999)
have shown that for full-field isoluminant stimuli intentional luminance
contamination does not appreciably alter the chromatic response. Our results,
demonstrating very small peripheral luminance response at low spatial
frequencies (see Figure 6), support this
observation. Therefore, luminance artifacts in larger field sizes would not
contribute appreciably to the overall response.
Luminance mfVEPs show reliable response contributions
from all field locations for a stimulated area up to 52 degrees ( Klistorner & Graham, 2000). Hood, Yu,
Zhang, Albrecht, Jaegle, & Sharpe ( 2002) were able to isolate responses from an
L-M-cone pathway in all field locations subtending 22.2 degree in radius using
the pattern-reversal mode. We do not believe that the different modes of
stimulus presentation (pattern-reversal versus onset) is the basis for low
peripheral responses in our recordings.
It has been shown that isoluminant VEPs for tritan and
L-M axes are tuned to low spatial frequencies ( Rabin et al., 1994). The negative response
(C II) decreases in amplitude and increases in latency as spatial
frequency moves above about 3 cycles per degree (cpd). The small peripheral
responses to chromatic reversal stimuli might be due to high spatial frequency
attenuation of chromatic responses ( Murray et
al., 1987; Rabin et al., 1994). The
optimal stimuli to elicit robust chromatic pattern-onset full-field VEPs are
sinusoidal gratings with spatial frequencies 0.5 – 2 cpd depending on the
chromatic axis ( Murray et al., 1987; Rabin et al., 1994; Porciatti & Sartucci, 1999). It would be
preferable to use sinusoidal gratings (Gabor patches) to record chromatic onset
mfVEPs, but this type of stimulus is currently not available with the
VERIS TM system. In addition, the scaling of the stimulus pattern we
used produced smaller peripheral responses and should be re-examined in view of
the dependence of cortical signals on retinal eccentricity of the stimulus ( Celesia & Meredith, 1982; Yiannikas & Walsh, 1983) due to cortical
magnification ( Rovamo & Virsu, 1979; Horton & Hoyt, 1991). Further studies are
needed to optimize the scaling and spatial frequency for each chromatic axis.
In conclusion, mfVEPs to chromatic onset patterns
display characteristic features shown for full-field recording. The validity of
the multifocal approach was verified by experiments demonstrating S-cone
isolation and isoluminance at different field locations. Stimulation with
pattern-onsets produced reliable and consistent waveforms for chromatic
modulation, but pattern-reversal stimuli were superior for luminance modulation.
Further refinement of the methods introduced here for recording chromatic
multifocal pattern-onset VEPs have potential application for detecting early
changes in the retina and optic nerve associated with aging and disease.
This work was supported by grants from the National
Institute on Aging to Michael A. Crognale (AG1869-01) and John S. Werner
(AG04058), an NEI Core Grant (EY12576) and a Research to Prevent Blindness Jules
and Doris Stein Professorship (JSW). We thank Susan Garcia for her help with the
recordings. Commercial relationships: none.
1. In the following
sections we use the term mfVEP to refer to the pattern-onset mode unless
otherwise stated.
2. Further tests of
each condition would be needed to conduct a formal signal-to-noise analysis
(e.g. Zhang et al., 2002). We believe this will
be meaningful only after the stimulus and recording techniques are optimized.
3. When creating the
tilted axes stimuli at a luminance value of 37 cd . m -2,
we noticed that the VERIS TM system automatically makes slight changes
to some RGB values after the manual input. This apparently happens so that the
values comply with the luminance calibration tables implemented internally by
the VERIS TM system. We found that the RGB values remained largely
unchanged at the 30 cd . m -2 level. These changes had no
significant effect on the original three axes used in Experiment 1. In addition,
the transient tritanopia control experiment also suggests that the stimuli in
Experiment 1 were sufficiently close to the tritan and L-M axes.
4. The magnification
induced by the lens for this recording condition was 1.13 and had only a
negligible impact on the retinal area stimulated.
Adams, A. J., Rodic, R., Husted,
R., & Stamper, R. (1982). Spectral sensitivity and color discrimination
changes in glaucoma and glaucoma-suspect patients.
Investigative Ophthalmology and Visual
Science, 23, 517-524. [ PubMed]
Baseler, H. A., & Sutter,
E. E. (1997). M and P components of the VEP and their visual field
distribution. Vision Research, 37,
675-690. [ PubMed]
Baseler, H. A., Sutter, E. E.,
Klein, S. A., & Carney, T. (1994). The topography of visual evoked response
properties across the visual system.
Electroencephalography and Clinical
Neurophysiology, 90, 65-81. [ PubMed]
Berninger, T. A., Arden, G.
B., Hogg, C. R., & Frumkes, T. (1989). Separable evoked retinal and
cortical potentials from each major visual pathway: preliminary results.
British Journal of Ophthalmology, 73,
502-511. [ PubMed]
Brainard, D. H. (1997). The
Psychophysics Toolbox. Spatial Vision,
10, 433-436. [ PubMed]
Brainard, D. H., Peli, D. G.,
& Robson, T. (2002). Display characterization. In J. Hornak (Ed.),
The Encyclopedia of Imaging Science and
Technology, (pp. 172-188). New York: John Wiley and Sons.
Celesia, G. G., &
Meredith, J. T. (1982). Visual evoked responses and retinal eccentricity.
Annals of the New York Academy of Sciences,
388, 648-650. [ PubMed]
Crognale, M. A., Switkes, E., Rabin, J., Schneck, M.
E., Haegerstrom-Portnoy, G., & Adams, A. J. (1993). Application of the
spatiochromatic visual evoked potential to detection of congenital and acquired
color-vision deficiencies.
Journal of the
Optical Society of America A, 10, 1818-1825. [ PubMed]
Crognale M.A., Switkes E.,
Rabin J., Schneck M.E. Hægerström-Portnoy G. & Adams A.J. (1995)
Objective assessment of short wavelength sensitive (SWS) mechanisms with the
spatio-chromatic VEP: S-cone monochromacy and transient tritanopia. In:
Color Vision Deficiencies XII, Doc.
Ophthalmol. Proc. Ser. 57, Kluver Acad. Publ., Dordrecht,
pp-407-413.
Derrington, A. M.,
Krauskopf, J., & Lennie, P. (1984). Chromatic mechanisms in lateral
geniculate nucleus of macaque. Journal of
Physiology (London), 357, 241-265. [ PubMed]
Goldberg, I., Graham, S. L.,
& Klistorner, A. I. (2002). Multifocal objective perimetry in the detection
of glaucomatous field loss. American Journal
of Ophthalmology, 133, 29-39. [ PubMed]
Greenstein, V. C., Hood, D.
C., Ritch, R., Steinberger, D., & Carr, R. E. (1989). S (blue) cone pathway
vulnerability in retinitis pigmentosa, diabetes and glaucoma.
Investigative Ophthalmology and Visual
Science, 30, 1732-1737. [ PubMed]
Harding, G. F. A., Odom, J.
V., Spileers, W., & Spekreijse, H. (1996). Standard for visual evoked
potentials 1995. Vision Research, 23,
3567-3572. [ PubMed]
Hasegawa, S., & Abe, H.
(2001). Mapping of glaucomatous visual field defects by multifocal VEPs.
Investigative Ophthalmology and Visual
Science, 42, 3341-3348. [ PubMed]
Hood, D. C., Yu, A. L., Zhang,
X., Albrecht, J., Jaegle, H., & Sharpe, L. T. (2002). The multifocal visual
evoked potential and cone-isolating stimuli: implications for L- to M-cone
ratios and normalization. Journal of Vision,
2(2), 178-189, http://journalofvision.org/2/2/4/, DOI 10.1167/2.2.4.
[PubMed] [ Article]
Hood, D. C., Zhang, X.,
Greenstein, V. C., Kangovi, S., Odel, J. G., Liebmann, J. M., & Ritch, R.
(2000). An interocular comparison of the multifocal VEP: a possible technique
for detecting local damage to the optic nerve.
Investigative Ophthalmology and Visual
Science, 41, 1580-1587. [ PubMed]
Hood, D. C., Zhang, X., Hong,
J. E., & Chen, C. S. (2002). Quantifying the benefits of additional channels
of multifocal VEP recording. Documenta
Ophthalmologica, 104, 303-320. [ PubMed]
Horton, J. C., & Hoyt, W.
F. (1991). The representation of the visual field in human striate cortex.
Archives of Ophthalmology, 109,
816-824. [ PubMed]
Klistorner, A. I., &
Graham, S. L. (2000). Objective perimetry in glaucoma.
Ophthalmology, 107, 2283-2299. [ PubMed]
Klistorner, A. I., Graham,
S. L., Grigg, J. R., & Billson, F. A. (1998). Multifocal topographic visual
evoked potential: improving objective detection of local visual field defects.
Investigative Ophthalmology and Visual
Science, 39, 937-950. [ PubMed]
Kulikowski, J. J., Robson,
A. G., & McKeefry, D. J. (1996). Specificity and selectivity of chromatic
visual evoked potentials. Vision Research,
36, 3397-3401. [ PubMed]
MacLeod, D. I. A., &
Boynton, R. M. (1979). Chromaticity diagram showing cone excitation by stimuli
of equal luminance. Journal of the Optical
Society of America, 69, 1183-1186. [ PubMed]
Mollon, J. D., & Polden, P.
G. (1977). An anomaly in the response of the eye to light of short wavelengths.
Philosophical Transactions of the Royal
Society (London), B278, 207-240. [ PubMed]
Murray, I. J., Parry, N. R.,
Carden, D., & Kulikowski, J. J. (1987). Human visual evoked potentials to
chromatic and achromatic gratings. Clinical
Vision Sciences, 1, 231-244.
Pelli, D. G. (1997). The
VideoToolbox software for visual psychophysics: transforming numbers into
movies. Spatial Vision, 10, 437-442.
[ PubMed]
Porciatti, V., &
Sartucci, F. (1999). Normative data for onset VEPs to red-green and blue-yellow
chromatic contrast. Clinical Neurophysiology,
110, 772-781. [ PubMed]
Rabin, J., Switkes, E.,
Crognale, M., Schneck, M. E., & Adams, A. J. (1994). Visual evoked
potentials in three-dimensional color space - correlates of spatio-chromatic
processing. Vision Research, 34,
2657-2671. [ PubMed]
Rovamo, J., & Virsu, V.
(1979). An estimation and application of the human cortical magnification
factor. Experimental Brain Research,
37, 495-510. [ PubMed]
Sandberg, M. A., &
Berson, E. H. (1977). Blue and green cone mechanisms in retinitis pigmentosa.
Investigative Ophthalmology and Visual
Science, 16, 149-157. [ PubMed]
Smith, V., & Pokorny, J.
(1975). Spectral sensitivity of the foveal cone photopigments between 400 and
500 nm. Vision Research, 15, 161-171.
[ PubMed]
Sutter, E. E. (1991). The fast
m-transform: a fast computation of cross-correlations with binary m-sequences.
Society for Industrial and Applied Mahematics,
20, 686-694.
Sutter, E. E., & Tran, D.
(1992). The field topography of ERG components in man. 1. The photopic luminance
response. Vision Research, 32, 433-446.
[ PubMed]
Switkes, E., Crognale, M.,
Rabin, J., Schneck, M. E., & Adams, A. J. (1996). Reply to "Specificity and
selectivity of chromatic visual evoked potentials".
Vision Research, 36, 3403-3405. [ PubMed]
Switkes, E., & Crognale,
M. A. (1999). Comparison of color and luminance contrast: apples and oranges?
Vision Research, 39, 1823-1831. [ PubMed]
Westheimer, G. (1964).
Pupil size and visual resolution. Vision
Research, 4, 39-45. [ PubMed]
Yiannikas, C., & Walsh,
J. C. (1983). The variation of the pattern shift visual evoked response with the
size of the stimulus field.
Electroencephalography and Clinical
Neurophysiology, 55, 427-435. [ PubMed]
Yu, M. Z., & Brown, B. (1997).
Variation of topographic visually evoked potentials across the visual field.
Ophthalmic and Physiological Optics,
17, 25-31. [ PubMed]
Yu, M. Z., Brown, B., &
Edwards, M. H. (1998). Investigation of multifocal visual evoked potential in
anisometropic and esotropic amblyopes.
Investigative Ophthalmology and Visual
Science, 39, 2033-2040. [ PubMed]
Zhang, X., Hood, D. C., Chen, C.
S., & Hong, J. E. (2002). A signal-to-noise analysis of multifocal VEP
responses: an objective definition for poor records.
Documenta Ophthalmologica, 104,
287-302. [ PubMed]
|
|